🎈#Desflurane has a saturated vapour pressure of 664 mm of Hg @ 20 degree C and a boiling point of 22.6 degree C

🎈This is less than other agents; thus it requires a specific vaporizer

🎈The Tec Mk 6 vaporizer is a plenum vaporizer and is designed for desflurane.

🎈It operates at 39 °C and 2 atmospheres (All other vaporizers operate at atmospheric pressure and temperature). 

🎈It takes 5–10 minutes to warm up to the correct temperature.

🎈It will not operate until the vaporisation chamber has reached 39 °C.

🎈It needs an external power supply, and it has a 9 V battery as a backup in case of mains failure.

🎈Unlike other plenum vaporisers, the Tec Mk 6 vaporizer is designed so that the fresh gas flow does not enter the chamber. The desflurane vapour is added to the fresh gas flow as it leaves the vaporiser.

#anesthesia , #anaesthesia , #anesthetist

Reference: Al-Shaikh B, Stacey S. Essentials of Anaesthetic Equipment, 2nd edn. Edinburgh: Churchill Livingstone, 2002